The Precision LCII is Radiant's answer to a growing need for affordable characterization equipment.The Precision LCII is an ideal general purpose tester with a broad test range for thin films and bulk ceramics.The Precision LCII tester has a frequency rating of 5KHz at +/-200V built-in to the system. The Precision LCII Test System makes testing of thin films and bulk ceramics a fast and simple process.
The Precision LCII executes hysteresis, pulse, leakage, IV, and CV measurements without changing sample connections.With the addition of extra fixtures, the Precision LCII can measure pyroelectric properties, magnetoelectric properties, transistor characteristics, cryogenic properties, bulk and/or thin film piezoelectric properties.
The Precision LCII II is offered with a variety of internal amplifiers.The Precision LCII is offered in a ±10V, 30V, 100V, 200V, and 500V built-in drive volt option. The Precision LCII can be expanded to 10kV with the addition of a high voltage interface and amplifier.Includes Vision Data Acquisition and Management System Software.
Tester Parameter | LC II |
Voltage Range (built-in drive voltage) | ±10V, ±30V, ±100V, ±200V or ±500V built-in |
Voltage Range with an external amplifier and high voltage interface (HVI) | 10KV |
Number of ADC Bits | 18 |
Minimum Charge Resolution | <10fC |
Minimum Area Resolution (assuming 1 ADC bit = 1μC/cm2) | <1μ2 |
Maximum Charge Resolution | 276μC |
Maximum Area Resolution (assuming saturation polarization = 100μC/cm2) | 2.76cm2 |
Maximum Charge Resolution with High Voltage Interface (HVI) | 27.6mC |
Maximum Area Resolution (assuming saturation polarization = 100μC/cm2) w/o HVI | >100cm2 |
Maximum Hysteresis Frequency | 5KHz @ 10V |
5KHz @ 30V | |
5KHz @ 100V | |
5KHz @ 200V | |
2KHz @ 500V | |
Minimum Hysteresis Frequency | 0.03Hz |
Minimum Pulse Width | 50μs |
Minimum Pulse Rise Time (5V) | 40μs |
Maximum Pulse Width | 1s |
Maximum Delay between Pulses | 40ks |
Internal Clock | 25ns |
Minimum Leakage Current (assuming max current integration period = 1 seconds) | 1pA |
Maximum Small Signal Cap Frequency | 20KHz |
Minimum Small Signal Cap Frequency | 1Hz |
Output Rise Time Control | 103 scaling |
Input Capacitance | -6fF |
Electrometer Input All Test Frequencies for all test at any speed | Yes |