Precision LCR Meter

  • Product No:GenRad 1693
  • Manufacturer:IET LABS, Inc. 

Description:

The GenRad 1693 RLC DIgibridge provides the best combination of features designed to optimize productivity in all testing environments. This bridge features highly accurate measurements of 11 impedance parameters and testing versatility through selection of a wide-range of test frequencies, speeds, and voltages. Its easy-to-read, dual display shows both primary and secondary measurement parameters.

The 1693 RLC Digibridge is controlled by a microprocessor, allowing for many automated functions including:

• testing

• parameter selection

• test frequency and voltage selection

• limit comparison

• binning

• zeroing

The automated capabilities of this meter can be further extended with the addition of an optional IEEE-488 interface, which allows for remote operation, programming, and data acquisition.

The combination of its high-accuracy, wide range of impedance parameters and test conditions, and automated functions make the 1693 Digibridge ideal for production testing, component design and evaluation, process monitoring, and dielectric measurements.


Features:

• The world's de facto standard for ac resistance, low-frequency inductance, and capacitance measurement

• Used as test standard at National Labs

• 0.02% accuracy for R,L,C, G, Z, and Y

• 0.0001 accuracy for Dissipation and Q

• 11 Impedance Parameters

• Programmable test frequencies from 12 Hz to 200 kHz for testing versatility

• Programmable test voltages from 5 mV to 1.275 V

• Dual display featuring 5-digit readout for RLC and 4-digit readout for D and Q

• Extremely reliable: over 30 years of history

• Optional IEEE-488 interface available


  • Measurement parameters: R/Q, L/Q, C/D (series or parallel), R/X (series), G/B (parallel), Z/Angle or Y/Angle

  • Parameter selection: Auto parameter (RLC) with manual selection

  • Accuracy: Basic RLCGZY: ±0.02%

                     Basic QD: ±0.0002 (±0.0001 in PPM mode)

                     Basic RXB: ±0.02%

                     Θ: ±0.01°

  • Test frequencies:

       Range: Over 500 selectable test frequencies ranging from 12 Hz to 200 kHz

       Accuracy: 0.01%

  • Binning: Pass bins: 13 pass bins for RLCGZY

                  Fail bins: 2 fail bins, RLCGZYΘ

  • Sorting capabilities: Bin number, Delta RLC, Delta %, Value

  • Applied voltage: 5 mV to 1.275 V (programmable in 5 mV steps)

  • Bias: Internal: 2.0 Vdc

             External: 60 Vdc max

  • Range selection: Autoranging with manual hold

  • Measurement mode: Continuous or triggered, with averaging of up to 256 measurements

  • Measurement speed: Up to 19 measurements per second

  • Display format: Dual display featuring 5 full digit LED for RLCGZY and  full digit LED for DQRXBΘ

                                      Automatically positioned decimal points and minus signs where appropriate

                                      Individual LED indicators for parameters, and measurement units

  • Interfaces: Optional IEEE-488.2 with updated SCPI commands

  • General features:

       Charged capacitor protection (1 joule)

       Keyboard lock (protects test conditions)

       Constant voltage mode (25 Ω source)

       Programmed delay (1 to 99,999 ms)

       DQ in ppm

       Bin count summary

       Programmed integration time

       Median value

  • Power: 90-250 Vac    50-60 Hz    60 W max

  • Environmental conditions: Operating conditions: 0° to +50°C, <85% RH

                                               Storage conditions: -45°C to +75°C