Pioneering Ferroelectric Test System

  • Product No:Multiferroic II
  • Manufacturer:Radiant Technologies

The Precision Multiferroic II Ferroelectric tester is the most advanced test system on the market.The Multiferroic II has a unique frequency rating of 270KHz at +/-100V built-in to the system.The Multiferroic II tester makes testing of  thin films and bulk ceramics a fast and simple process.  

 

The Multiferroic II executes hysteresis, pulse, leakage, IV, and CV measurements without changing sample connections.With the addition of extra fixtures,the Multiferroic II can measure pyroelectric properties, magnetoelectric properties, transistor characteristics, cryogenic properties, and bulk and/or thin film piezoelectric properties.  

  

The Precision Multiferroic II is offered with a variety of internal amplifiers.The Multiferroic II is offered with a  ±100V, 200V and 500V built-in drive volt option.The Multiferroic II can be expanded to 10kV with the addition of a high voltage interface and an amplifier.Includes Vision Data Acquisition and Management System Software. 


Tester Parameter

MultiFerroic II

Voltage Range (built-in drive voltage)±10V, ±30V, ±100V, ±200V or ±500V built-in
Voltage Range with an external amplifier and high voltage interface (HVI)

10KV

Number of ADC Bits

18

Minimum Charge Resolution

0.80fC

Minimum Area Resolution (assuming 1 ADC bit = 1μC/cm2)

0.080μ2

Maximum Charge Resolution

5.26mC

Maximum Area Resolution (assuming saturation polarization = 100μC/cm2)

52.6cm2

Maximum Charge Resolution with High Voltage Interface (HVI)

526mC

Maximum Area Resolution (assuming saturation polarization = 100μC/cm2) w/o HVI

>100cm2

Maximum Hysteresis Frequency

270KHz @ 10V


270KHz @ 30V


270KHz @ 100V


100KHz @ 200V


5KHz @ 500V

Minimum Hysteresis Frequency

0.03Hz

Minimum Pulse Width

0.5μs

Minimum Pulse Rise Time (5V)

400ns

Maximum Pulse Width

1s

Maximum Delay between Pulses

40ks

Internal Clock

25ns

Minimum Leakage Current (assuming max current integration period = 1 seconds)

1pA

Maximum Small Signal Cap Frequency

1MHz

Minimum Small Signal Cap Frequency

1Hz

Output Rise Time Control

10scaling

Input Capacitance

-6fF

Electrometer Input All Test Frequencies for all test at any speed

Yes