The industry leading 3200A series provides by far the fastest and safest solution for calibration of Loop, RCD, PAT's and insulation testers.
Traditional testing practices require many pieces of test equipment and exposure to hazardous voltages, with a typical multifunction tester taking up to 45 minutes to test.
Using the 3200A, calibration time can be dramatically reduced to under 15 minutes, all while increasing safety. Cost effective calibration can be put into practice easily, providing a fast return on investment.
A dedicated localized mains socket on the front panel provides direct connect for Loop and RCD testers, removing the need to make connections to the mains/line voltage using 4mm type test leads and crocodile clips. This important safety feature eliminates the risk of shock to the engineer as connections are made with standard mains/line plugs.
CONTINUITY TESTING
Continuity Resistance
Continuity Variable Range: 0.2Ω to 20Ω (0.01Ω Resolution)
Continuity Fixed Outputs : 100Ω : 1kΩ
Standard Accuracy : 1% ± 25mΩ
High Accuracy (Option) : 0.25% ± 25mΩ
Continuity Current Measurement (Into 1Ω)
Range : 0 to 320mA (0.1mA Resolution)
Accuracy : 1.3% ± 6 Counts
INSULATION TESTING
Insulation Resistance
Range : 10kΩ to 2GΩ (10GΩ Option : 5%)
Accuracy : 10kΩ to 5MΩ 0.3% (0.1% Option)
5MΩ to 2GΩ 3% (1% Option)
Resolution : 10kΩ
Max Volt/Power : 1.1kV or 1 WATT (15kV Option)
Insulation Test Voltage Measurement
Ranges : 50V, 100V, 250V, 500V, 1kV
(0.1V Resolution)
Current Measurement : 0.5mA, 1mA
Accuracy : 1% ± 8 Counts
AC Voltage Output
Range : 100V, 200V, 230V, 300V, 400V
Accuracy : 0.2% ± 1 Counts
RCD TESTING
Time Range : 20ms to 5s (10ms Resolution)
Accuracy : ± 0.7ms
Current Range : 3mA to 3A
Resolution : 0.01mA (to 200mA)
: 0.1mA (to 1A) • 1mA (to 3A)
Accuracy : < 190ms 5% ± 20 Counts
: up to 5s 1.2% ± 6 Counts
Timing Modes : Immediate & Zero Crossing
Current Modes : I, ½ I, 2I, & 5I
Display Modes : 0° and 180° Phase
Duration Range : 10ms to 5s
Resolution : 0.1ms
LOOP TESTING
Resistances
0.05Ω, 0.1Ω, 0.22Ω, 0.33Ω, 0.5Ω, 1Ω, 5Ω, 10Ω, 100Ω, 1kΩ
Resolution : 0.1mΩ(to 10Ω) ; 1mΩ (to 1kΩ)
Accuracy : 0.5% ± 4mΩ
Auto Loop (Option)
Maximum Correction : 2Ω
Resolution : 100μΩ
Accuracy : ±18mΩ
Measurement Current : 4A
PAT TESTING
Earth Bond Resistance
Values : See Loop Values
Accuracy : 0.5% ± 4mΩ
Resolution : 0.1mΩ (to 10Ω) : 1mΩ (to 1kΩ)
Earth Bond Current Measurements
Ranges : 100mA, 10A, 30A
Resolution : 1mA, 10mA, 10mA
Accuracy : 1.5% ± 6 Counts
PAT Insulation Resistance
Range : 30kΩ to 2GΩ (10kΩ Resolution)
Accuracy : See Insulation Resistance
PAT Leakage
Ranges : 2mA, 4.7mA, 7.7mA (1uA Resolution)
Accuracy : 1.5% ± 2 Counts
PAT Leakage Test Voltage
Range : 100V to 300V RMS (0.1V Resolution)
Accuracy : 1.5% ± 9 Counts
Load Testing
Ranges : S/C, O/C & 0.13kVA
Accuracy : 440Ω ± 5%
PAT TESTING : FLASH (OPTION)
Flash Test Voltage
Class 1 : 1kV to 1.8kV (1V Resolution)
Load Resistance : 600kΩ (2.5mA @ 1.5kV)
Class 2 : 2kV to 3.6kV (1V Resolution)
Load Resistance : 1.2MΩ (2.5mA @ 3kV)
Accuracy (Class 1&2) : 4% ± 10 Counts
Flash Current Measurement
Range : 1mA to 3mA
Resolution : 10uA
Accuracy : 5%
BREAKDOWN TESTING (OPTION)
Voltage Range : 3kV • 12kV
Resolution : 10V
Accuracy : 0.5% ± 3 Counts
Current Range : 200uA • 2mA • 20mA
Resolution : 100nA • 1uA • 10uA
Accuracy : 0.5% ± 4 Counts (200uA)
LINE VOLTAGE MEASUREMENT
Range : 200V to 260V
Resolution : 0.1V
Accuracy : 0.8% ± 6 Counts
GENERAL SPECIFICATIONS
Line Power : 230V/110V 50Hz /60Hz
Dimensions : 45 x 44 x 14cm • Weight : 10kgs
Warm-up Time : Double the time last used to 20 minutes max.
Interface : RS232 • USB (Via Adapter)