Decade Voltage Divider

  • Product No:1455 series
  • Manufacturer:IET LABS, Inc. 

The GR 1455 Kelvin-Varley Voltage Dividers provide accurately known voltage ratios from 0.00001 to 1.00000 for use in many common measurements:

  • 1. Voltage gain or attenuation
  • 2. Linearity of potentiometers and other controls
  • 3. Frequency response of audio and rf networks
  • 4. Transformer turns ratio
  • 5. Voltmeter calibration

A resistive divider of the Kelvin-Varley type, the 1455 has precision resistors throughout (rather than in selected positions only) for over-all high accuracy. Linearity is as low as 0.02 ppm of input.


Match your needs exactly. Select input impedance. voltage rating, frequency range, 4- or 5-d ial resolution, bench or rack mounting.


Temperature Coefficient: <20 ppm for each resistor. Since voltage ratios are determined by resislors of simi lar conslruc lion, nel ambient lemperature effecls are very small.


BASIC SPECIFICATIONS



1455-A1455-B1455-BH
Number of Dials: 455
Input Resistance: 10 kΩ10 kΩ100 kΩ
Accuracy of Input R: (ppm)150150150
Input Voltage Rating: 230 V 230 V 700 V 
Frequency Response: 850 kHz 690 kHz 69 kHz 
Resolution: (ppm of input) 1001010
Linearity (sum of A & B)


A. Absolute Linearity


0.00001 to 0.00010n/a0.030.02
0.00010 to 0.001000.30.30.2
0.00100 to 0.01000232
0.01000 to 0.10000151010
0.10000 to 1.00000302020
B. Terminal Linearity (ppm of input).


FOUR-TERMINAL (low output terminal)0.040.040.004
THREE-TERMINAL0.20.20.02